5 – 33Operation Manual KjelDigester K-446/K-449
5 Installation
5.7 Condensate bottle, installation
CAUTION
Risk of acid or peroxide fume inhalation!
Minor or moderate chemical burns of the airways.
1. Keep the suction hoses between the KjelDigester and Scrubber as straight
as possible to prevent liquid blocking the hoses.
2. Make sure the condensate bottle is placed lower than the suction module.
3. Operate and maintain the Scrubber according to its operation manual.
4. Make sure the hoses are not damaged.
1. Connect the hosec with the hose connector to the fume collection tubea.
2. Secure the connection with the catchpot clipb.
3. Connect the hosec coming from the KjelDigester
B
to the condensate bottleb.
∙ 025100 Condensate bottle
4. Connect the hosea from the Scrubber
A
to the condensate bottle.
5. Close the 2 remaining connections of the condensate bottle with 2plugs.