5 Putting into operation
20 Pump Combinations Operation Manual, Version E
1
3
42
Connect the pump module a to the pump control-
ler b via data bus RS-485 by means of the cable
RJ45 d.
c TTL IN/OUT.
Fig. 5.6: Rear connections of pump controller and pump module
The overpressure sensor switches off the pump if the pressure exceeds 10 bar. This maximum pres-
sure of 10 bar cannot be changed. The installation of the overpressure sensor is easy to do and
explained in the following:
Overpressure sensor
Clip the overpressure sensor onto the Pump Module
C-601.
Connect the overpressure sensor to the correspon-
ding connector TTL IN/OUT at the back of the Pump
Controller C-610.
Fig. 5.7: Installing the overpressure sensor