4 Description of function
24 V-700/710 Operation Manual, Version E
4.4 Vacuum Module V-802 Lab Vac
The Vacuum Pump V-700/710 can be used together with the Vacuum Module V-802 LabVac as a
laboratory vacuum system. If the pump is used without a Vacuum Module V-802 as a laboratory
vacuum system it will automatically reduce the suction power after one hour of operation and start
working in the ECO
2
Mode (see also chapter 6.1.5).
1
2
3
4
Fig. 4.11: Front view LabVac
NOTE
The back view of the LabVac as the same as for the EasyVac, see Fig. 4.8.
1
2
a Pump not running b Pump running
Fig. 4.12: Pressure course in LabVac mode
a Start
By pressing the START button the pump starts evacuating the system. When a constant vacuum is
reached the pump stops. When the pressure rises over the hysteresis at 50 mbar the pump starts to
operate again until a constant vacuum is reached.
Continuous
By pressing the START button for more than 2 seconds the pump enters the Continuous mode,
where the pump runs constantly.
b Stop
When pressing the STOP button the pump stops and the system is aerated by means of the valve at
the controller.