4 Putting into Operation
Fig. 4.6: Reflux glass assembly
4.7 Installation of the Reflux Glass Assembly
All glassware used must be intact, with no signs of cracks,
spreading impact marks, or other damage. Inspect the glass-
ware visually before installing it.
s &IXTHEREmUXDISTRIBUTIONHEAD in position on the gear
head using a DN 70 EasyClamp connection.
s &ITTHECOOLINGWATERHOSENIPPLES to both coolers .
s )NTRODUCEBOTHCOOLERS, together with the cooler hol-
ders , into the pivoting clamps and connect them to
the distribution head (2 x DN 40 EasyClamp connections).
s !LIGNTHECOOLERS in a vertical position and fix in place
with the pivoting clamps .
s 3CREWTHESHUTOFFTAP into the distribution head
and tighten it firmly.
s )NSERTTHEINLETTAP into the distribution head and tighten
the SVL 30 nut.
s 0LACE THE CONDENSATE COOLER on the drip tray and
connect it using the EasyClamp connection. Fit the con-
necting pipe between the condensate cooler and the
distribution head and secure it (2 x DN 25 EasyClamp
connections).
s 3CREW THE TEMPERATURE SENSOR into the distribution
head .
s 4IGHTENALL%ASY#LAMPCONNECTIONSEVENLYANDINPARALLEL
s 4URNTHEPOSITIONINGNUTTORAISETHEmASKSUPPORT#ONNECT
the receiving flask to the condensation flask.