6 Operation
30 K-425 / K-436 Operation Manual, Version A
1
2
Rear side
Front side
Insert the prepared sample tubes into the rack(s) start with the first (front) position. •
When the digestion session does not require a full set of samples, then fit unused positions with a •
glass cap to seal the suction circuit!
Note
Make sure the adapter piece is tightened well without damaging the glassware.•
Keep suction hose(s) as short as possible under the given circumstances.•
Unused positions should be arranged at the rear end of the rack.•
3
3
4
Rear side
Lift the metal clips to unlock and shift the black handles • 3 aside.
Carefully push down the suction module • 4 straight onto the sample tubes and shift the handles
back into locking position.
The metal clips must snap in to securely lock the handles position!