5 Putting into operation
22 V-850/855 Operation Manual, Version E
5.3 Installing the AutoDest probe
1
To install the AutoDest probe proceed as follows:
Insert the AutoDest probe into the opening at the top of the condenser.
Push the probe down between the inner condenser coil and the vacuum tube in the center of the
condenser.
Position the probe as shown in Fig. 5.3. Make sure that the prope tip position corresponds to posi-
tion a in Fig. 5.3.
Tightened screw cap at the top of the condenser.
Plug the cable of the AutoDest probe into the AS/SB socket of the controller (position b in the
figure above).
NOTE:
Calibrate the AutoDest probe, see also chapter 6.7.5.