8 Troubleshooting
149
K-375/376/377 Operation Manual, Version B
15’114 Sampler: Error during EEPROM writing.
Adjustment value not saved
Call service
17’001 Steam generator overtemperature Switch distillation unit off and on or call service
17’002 Water level not reached Check steam generator water supply or call service
18’001 Stirrer out of order Check stirrer cable or exchange stirrer
50’001 Device powered on System message, no error.
50’002 Device powered off System message, no error.
50’003 User login System message, no error.
50’004 User logout System message, no error.
50’005 Verify data consistency time System message, no error.
50‘006 Device power failure during determination System message, no error.
50’007 Automatic export error System message, no error.
50‘008 Sequence information System message, no error.