BÜCHILabortechnikAG Installation|5
PrepChromC700 39/110
5 Installation
5.1 Modules,assembly
Fig.15:Assemblyorder
1 Fractioncollectormodule 2 Pumpmodule
3 Controlunit 4 Valvemodule
5 UVDetectormodule
u Selectasuitablelocationbyensuringadequateairspaceof15cmaround
thesystemtoprovideventilationofthecontroller,thepumpandthedetec
torcoolingfans.
Totalweight:53kg
NOTICE
Riskofspilledliquids.
LiquidsormaterialplacedontopoftheUVDetectormodulecanfalldown.
u DonotplaceanyliquidsormaterialontopoftheUVDetectormodule
withouttheoptionalsolventtray.
u Placethesinglemodulesaccordingtotheassemblyorder.
u Removethecableties(1)and(3)fromthefractioncollectorarm(2).