3 Technical data
26 Multivapor™ Operation Manual, Version E
3.3 Technical data overview
Table 3-11: Technical data
Dimensions ( W×H×D) 270×400×400 mm
Weight P-6: 22 kg, P-12: 21 kg
Connection voltage 100 – 120 or 220 – 240 VAC ± 10%
Fuse T 3.1 A L 250 V (220–240 V)
T 6.3 A L 250 V (100–120 V)
Power consumption max. 800 W
Mains connection 3-pole (P, N, E) via power cord
Frequency 50/60 Hz
Installation category II
Degree of protection IP21
Pollution degree 2
Rotation speed range P-6: 0– 370 rpm, P-12: 0 – 485 rpm
Temperature control range 20 – 95 °C
Temperature accuracy ± 3 °C (instrument calibrated at 20 °C)
Display Set and actual temperature
Max. size of sample tube P-6: OD 16 – 60 mm, L = 110 – 150 mm;
P-12: OD 15 – 30 mm, L = 15 – 150 mm
Max content of sample tube P-6: 160 mL, P-12: 30 mL
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
5 – 40 °C
up to 2000 m
maximum relative humidity 80% for temperatures up to
31 °C, and then linearly decreasing to 50% at 40 °C
Temperature resistance P+G ca. -70 °C – 60 °C
Temperature resistance
P+G low temperature
-80 °C – 50 °C
Temperature resistance protective shield < 160 °C