BÜCHILabortechnikAG Productdescription|3
OperationManualRotavapor®R100 19/60
3.5 Safetyfeatures
3.5.1 Overheatcutout
Toprotectagainstoverheating,theheatingbathhasathermostaticbathtemperature
control.
Inadditionitisfittedwithanelectronicandamechanicaloverheatcutout.
Theelectronicoverheatcutoutmonitorsthetemperaturelimit(theactualtemperature
ofthebathmustnotexceedthespecifiedtemperaturebymorethan2°Cforlonger
than2minutes)theheatingrate(theactualtemperaturemustnotincreasebymore
than5°Cwithin5seconds)andthefunctionofthetemperaturesensor.Iftheelec
tronicoverheatcutouttrips,thedeviceshouldbeinspectedbyaBUCHIservicetech
nician.
Themechanicaloverheatcutoutconsistsofabimetallicthermostatthatimmediately
cutsoffthepowersupplyathightemperatures(above145°C).Themechanicalover
heatcutouthastobemanuallyresetoncetheheatingbathhascooleddown(see
Chapter8.2.1"Resettingoverheatsafetycutout",page44).
3.5.2 Protectionagainstelectricaloverload
Theheatingbathisequippedwithfusestoprotectagainstelectricaloverload.
Therotarydriveunithasanelectricaloverloadcutout.
3.5.3 Clipsandholders
Combiclipsforfixingtheevaporatingflaskandsafereleaseofstickingground
glassjoints
Balljointclampsforsecurelyfixingthereceivingflask
Laboratorystandandholderforfixingglassapparatus
Capnutforfixingthecondenser
3.5.4 Glassware
Highqualityborosilicateglass3.3
GL14hosebarbsforpreventingglassbreakage
3.5.5 Optionalaccessories
Safetyguardtoprotectuseragainstglasssplinters,solventsplashesorhotheating
fluidintheeventofaccidentsorimplosions
Laboratorystandforprovidingadditionalfixingforcondenser
(seeChapter10.2"Sparepartsandaccessories",page48)