8|Helpwithfaults BÜCHILabortechnikAG
42/60 OperationManualRotavapor®R100
8 Helpwithfaults
8.1 Faults,possiblecausesandremedies
Fault Possiblecause Remedy
Devicedoesnot
work
Deviceisnotconnectedto
powersupply
u Checkpowersupply(see
Chapter5.6"Electricalcon
nections",page25).
Fuseblown u Replacefuse(seeChapter
8.2.2"Replacingthefuse",
page45).
u Iffaultrecurs,contact
BUCHICustomerService.
Heatingbath
failstoheatup
Deviceisnotconnectedto
powersupply
u Checkpowersupply(see
Chapter5.6"Electricalcon
nections",page25).
Hightemperaturecutouthas
tripped
u Resettheoverheatsafety
cutout(seeChapter8.2.1
"Resettingoverheatsafety
cutout",page44).
Fuseblown u Replacefuse(seeChapter
8.2.2"Replacingthefuse",
page45).
u Iffaultrecurs,contact
BUCHICustomerService.
Systemleaking Groundglassjointsnotgreased u Greasegroundglassjoints.
Tubingleaking u Replacetubing(seeChap
ter5.5"Connectingthevac
uumandcoolingmediatub
ing",page24).
Sealdefective u Replaceseal(seeChapter
7.3"Checkingseals",page
40).
Vacuumisnot
achieved
Backevaporationfromthere
ceivingflask
u Emptythereceivingflask
(seeChapter6.3.2"Remov
ingthereceivingflask",
page38).
Temperaturedifferencebe
tweenevaporatingflaskand
condenserlessthan20°C
u Increasecooling(seeChap
ter6.2.2"Adjustingdistilla
tionconditions",page34).
Systemleaking u Checksystemforleaks(see
Chapter7.2"Checkingthe
systemforleaks",page39).
Waterpressureofglassfilter
pumptoolow
u Increasewaterflow(see
pumpoperatinginstructions).
Vacuumpumptooweak u Usesuitablydimensioned
vacuumpump.