5.4 Fittingtheglassstopcock
Materialsrequired:
GreaseGlisseal40blue(ordernumber048197)
u Connectasupplytubingforsolvent(2).
u ApplyathincoatofGreaseGlisseal40bluetotheglassstopcock(1).
u Insertglassstopcockintothecondenserandturntothedesiredposition.
Theglassstopcockperformsvariousfunctionsdependingonitsposition:
Glassstopcockhandlepointingforwardsorbackwards:thesystemissealed.This
isthestandardpositionfordistillation.
Glassstopcockhandlepointingupwards:systemisbeingvented.
Glassstopcockhandlepointingdownwards:additionalsolventcanbefedintothe
evaporatingflaskviaaconnectedfeedingtubing.