Operation Manual D-480 ©BÜCH Labortechnik AG
8 - Maintaining the instrument 181
If deleting has not had the desired effect, you can undo it. To do so, select the deleted sample
and click the "Include/Exclude" button again. The sample is now included again.
Once you have excluded all samples with significant deviations from the calibration curve, click
the "Next" button. The next calibration curve appears in the calibration view.
Once you have optimized the last calibration curve and clicked the "Next" button, the
"Calibration coefficients" dialog is displayed.
Check...
... the calibration coefficients "a" to "e".
Note:
For the lower calibration range it is important that coefficient "a" is 1x10
-3
... 3x10
-2
or less.
Coefficient "a" corresponds to a calibrated blank value and should be as small as possible.
... the possible working range defined in "Min." and "Max.".
Decide:
If...
... you accept the calibration coefficients: Click OK. The calibration coefficients are applied.
... you do not accept the calibration coefficients: Click Cancel. Calibration is aborted.