Operation Manual D-480 ©BÜCH Labortechnik AG
10 - Menu and dialog descriptions 309
Name Textbox/label Field to input the name of the standard sample.
{Element} conc. [%]
(or {Element} conc.
[ppm])
Textbox/label Field to input the theoretical percentage content of the
corresponding element in the respective sample.
{Element} tolerance
[%]
Textbox/label Field to input the tolerance limit of the corresponding
element. The tolerance limit is required if standard
tolerance monitoring is enabled.
New
Button Allows the definition of a new standard sample. This
function empties the input boxes, thus data for a new
standard sample can be entered.
Delete
Button Deletes a standard sample previously selected in the
listbox.
Save/Create
Button Saves changes of an existing standard sample or creates
and saves the settings of a newly defined standard sample.
Close
Button Closes the dialog.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Defining standard substances as factor and monitor samples (on page 122).
Defining standard substances as calibration samples (on page 122).
Viewing list of defined factor, monitor and standard samples (on page 121).
Keywords (Dialog)
Function
In this dialog you specify keywords for conditioning samples and blank value samples. A key word is
a reserved sample name that the software recognizes and interprets in a special way.
Note
Keywords are already defined for conditioning samples and blank value samples. You can change
keywords, but you must observe the following:
The names must be distinguishable.
The program does not distinguish between uppercase and lowercase, so "BLANK" has the same
meaning as "blank", for example.
Open
Open the dialog via Options > Settings > Keywords.
Elements
The following describes the elements of the dialog.
Conditioning sample
name
Textbox Field to input the key word for conditioning samples.
Name blank value Textbox Field for input of the keyword for blank samples
OK
Button Saves the keywords.
Cancel
Button Discards keywords not yet saved.