Operation Manual D-480 ©BÜCH Labortechnik AG
10 - Menu and dialog descriptions 311
Open
Open the dialog via Options > Settings > Calculation.
Elements
The following describes the elements of the dialog.
In section "Factor determination":
sequentially Radiobutton The daily factor is computed sequentially.
total Radiobutton The daily factor is computed as a total.
manual input Radiobutton The daily factor can be entered manually, there will be no
calculation of the daily factor.
autom. factor
calculation
Checkbox calculation of the daily factor starts automatically after the
end of every analysis.
separate upper/lower
calibration range
Checkbox Determines whether valid factors shall be computed for
both calibration ranges or separate ones for each range.
In section "Blank value determination":
sequentially Radiobutton The blank value is computed sequentially.
total Radiobutton The blank value is computed as a total.
manual input Radiobutton The blank value can be entered manually, there will be no
calculation of the blank value.
autom. blank value
calculation
Checkbox calculation of the blank value starts automatically after the
end of every analysis.
Remaining elements:
OK
Button Saves new settings
Cancel
Button Discards settings not yet saved.
Further information
Further information includes instructions in which the dialog is used and knowledge articles
concerning the dialog:
Determining the daily factor (on page 145).
Determining blank values (on page 144).
Specifying the computation method for blank value and daily factor (on page 123).
LIMS settings (Dialog)
Function
This dialog is used to make various adjustments to a laboratory information and management system
(LIMS). The dialog has the following tabs:
On the "LIMS data" tab you select which data to send to the LIMS and whether they should be
sent automatically after every analysis.
On the "Communication" tab you specify whether and how to transfer data.