5 Putting into operation
57 R-220 SE Operation Manual, Version F
Basic heating transfer medium requirements
Do not use media which are corrosive against the heating bath and its subcomponents or the
evaporating flask. E.g. add a sufficient amount of borax when deionized water is used!
The medium must be liquid at room temperature.
Do not mix liquids of unknown composition!
!
DANGER
Death or serious burns by use with highly flammable substances in the heating bath.
Do not use heating transfer medium with a flash point below 200 °C
At liquid exchange, the heating bath must be clean and dry
!
DANGER
Death or serious burns by deflagration of hot heating bath liquids.
Do not mix or add water to other heating transfer media with a higher boiling point
The replenishment valve must not be connected to a water source when another heating
transfer medium than water is used
NOTE
Do not switch on the heating when the heating transfer medium level is unsufficiently low for
evaporation tasks.
Drain and clean the bath regularly to maintain the system in good working order.
Ambient humidity can diffuse into the heating medium and condensated water can gather at the
bottom when the bath has not been used over a longer period. To drive out unwanted water let
the evaporating flask rotate and heat up the bath above 100 °C for some minutes.