8 Troubleshooting
84 R-220 SE Operation Manual, Version F
Error Addition error indicator on display Possible cause Remedy
E50 ‘RECORD’ message flashes. No USB pendrive inserted or recog-
nized.
Insert valid pendrive.
Alternatively: Press “Record” button
to disable recording function.
Call service if problem is persistent.
E51 n.a. No free memory on USB pendrive. Empty pendrive.
Call service if problem is persistent.
E52 n.a. Write error on USB pendrive. Exchange pendrive.
Call service if problem is persistent.
E53 n.a. Inserted USB pendrive is write
protected.
Disable write protection.
Alternatively: Exchange pendrive.
Call service if problem is persistent.
E54 n.a. Inserted USB pendrive is write
protected.
Disable write protection.
Alternatively: Exchange pendrive.
Call service if problem is persistent.
Other error situations
In case the temperature reading at the display for vapor or cooling water is missing, check the
respective sensor cable and plug. Exchange a defective sensor and call service if the problem is
persistent.
8.3 Resettable automatic system fuses
Fuse triggered
Fuse ok
F1 — F4 (12A, 250V)
Resetting a system fuse
Switch off the Rotapavor
Push in the fuse for reset
Switch on the Rotavapor
NOTE
Device fuses may be triggered occasionally due to voltage peaks under high system load. When
fuses trigger frequently, inform your customer service! Do not try to repair the fuses!