7 Maintenance and repairs
77 R-220 SE Operation Manual, Version F
7.4 Distribution head seal and vacuum seal
Removal of the seal holder
a
b
c
a
Remove the snap flange coupling and the flask seal first. Pull out the vapor duct by hand.
b
Loosen the EasyClamp (DN70) and take off the distribution head (glass part). Subsequently,
completely remove the EasyClamp from the gearbox.
c
Pull out the seal holder by hand.
The distribution head seal can easily be pulled off by hand.
To remove the vacuum seal use the seal tool (order no. 020075). Hook it in under the
ring-shaped metal insert (!) and pull out the seal from the seal holder. Perform this action
in several minor steps and change the tool's position to avoid canting!